Performance evaluation of non-invasive cardiac output monitoring device (HemoVista) based on multi-channel thoracic impedance plethysmography technology
Jaehee Park, Byung-Moon Choi
Acute Crit Care. 2024;39(4):565-572.   Published online November 18, 2024
DOI: https://doi.org/10.4266/acc.2024.00731
  • 454 View
  • 79 Download
AbstractAbstract PDF
Background
A non-invasive method of measuring cardiac output (CO) can be beneficial in the care of critically ill patients. HemoVista (BiLab Co., Ltd.) is a medical device that measures CO non-invasively using multi-channel impedance plethysmography technology. The purpose of this study was to exploratively evaluate the performance of HemoVista in critically ill patients undergoing CO monitoring with the FloTrac (Edwards Lifesciences).
Methods
After non-invasively installing the HemoVista sensor in critically ill patients whose CO was monitored with the FloTrac, CO values measured by both devices were collected for 30 minutes. Cardiac output measured by both devices was selected every 10 seconds, creating approximately 360 data pairs per patient. Linear correlation analysis with Pearson correlation coefficients, Bland-Altman analysis, and four-quadrant plot analysis were performed to evaluate the performance of HemoVista.
Results
A total of 7,138 pairs of CO data from the 20 patients were included in the analysis. A significant correlation was observed between the two methods of measuring CO (Pearson's r=0.489, P<0.001). The mean bias was 1.03 L/min, the 95% CI for the limit of agreement was –1.83 L/min to 3.93 L/min and the percentage error was 55.8%. The concordance rate of time-dependent CO between the two devices was 14.6%.
Conclusions
It was observed that the current version of HemoVista has unsuitable performance for use in intensive care units. To be used for critically ill patients, the algorithm must be improved and reevaluated with an enhanced version.
Comparison of Impedance Cardiography with Thermodilution of Hemodynamic Parameters in Critically Ill Patients
Hyung Goo Kang, Sang Won Chung, Ki Hyun Byun, In Byung Kim
Korean J Crit Care Med. 2006;21(2):77-82.
  • 1,904 View
  • 17 Download
AbstractAbstract PDF
BACKGROUND
To evaluate the feasibility of noninvasive hemodynamic monitoring (impedance cardiography, ICG) in critically ill patients, we compared this technique with simultaneous invasive monitoring with a pulmonary artery thermodilution catheter.
METHODS
A prospective observational study was done comparing invasive monitoring and noninvasive monitoring in 12 critically ill patients. The cardiac output (CO), the stroke volume (SV) and the systemic vascular resistance (SVR) measured by using a standard thermodilution pulmonary artery catheter technique were compared with the corresponding measurements simultaneously using an ICG.
RESULTS
The value of CO, SV and SVR measured by ICG were closely correlated to those by the thermodilution methods [r: 0.659 (p<0.01), 0.536 (p<0.01), 0.738 (p<0.01)].
CONCLUSIONS
ICG can provide hemodynamic information previously available only by invasive monitoring with a thermodilution catheter.